Meditek Pass Thru Manual 2026

OPERATION & MAINTINENCE INSTRUCTIONS FOR

MDK18

Pass Thru Air Lock

Cleanroom Entry System

3

3

4

5

Table of Contents

Safety Warnings

Installation

Wall Mounted Pass Thru Installation

Fire Rated Pass Thru Installation

Operation & Maintenance

Interlock Operation

HEPA Filter Operation

Maintenance

Troubleshooting

Read and Save These Instructions

Please observe the following information related to your product

•

Read this guide before installing and/or operating the unit.

•

Keep this guide as a reference to the safety and operating instructions.

•

Observe warnings associated with handling, installing, using, and maintaining the MDK18.

•

Follow all instructions for set-up, operation, and use.

•

Operate ONLY from the type of power source indicated. If you are unsure of the type of power

available to you, contact your dealer or power company.

•

This unit may be equipped with a polarized alternating current (AC) plug with one blade wider

than the other. This plug will only fit into the power outlet one way for safety reasons. If the

plug does not fit, call an electrician to replace the outlet. DO NOT REMOVE THE ROUND

GROUNDING TERMINAL. THIS UNIT MUST BE GROUNDED FOR SAFETY REASONS AND FOR

PROPER OPERATION.

•

Do not place any power cords where they will have things placed on or against them, be

walked on, or be pinched, especially near the outlet, the plug, or where the power cord exits the

unit.

•

Overloading outlets and/or extension cords can result in fire and/or electrical shock.

•

ONLY use parts supplied by the manufacturer. Substitutions may result in fire, electrical shock,

or other safety hazards.

2

6

7

7

6

6

OPERATION & MAINTINENCE INSTRUCTIONS

MDK18

Pass Thru Air Lock Cleanroom Entry System

OPERATION & MAINTINENCE INSTRUCTIONS

MDK18

Pass Thru Air Lock Cleanroom Entry System

3

Safety Warnings

•

Never expose the assembly to moisture or rain.

•

If the unit gets wet, disconnect the power at the source and have an authorized service

inspection before using it again.

•

Do not pull on power cords to unplug the unit, and never handle the cord with wet hands.

•

Do not clean the unit with flammable chemicals.

•

Do not expose to explosive or hazardous vapors or materials.

•

Make sure the unit is well supported to prevent falling.

•

Do not block airflow to the unit.

•

Disconnect the power before servicing.

•

Do not operate below 0 degrees Fahrenheit or above 110 degrees Fahrenheit.

•

This unit is designed to operate in a non-hazardous (non-explosive) environment with non-

condensing air.

•

NEVER put objects into the blower.

•

Do not operate blower without metal blower guard.

•

Review application with your safety commissioner for proper use

Installation

Notes

•

Pass thrus must be installed in a flat and level wall capable of supporting the weight of the

pass thru. Weight depends on size, model, and options package.

•

The pass thru should be well supported on all sides and securely mounted so that it will not

shift when weight is inserted. If the pass thru is not securely mounted it may loosen over time.

•

Pass thrus need to be installed so that they are square and level. This is critical to ensure

proper alignment of the latch and interlock components.

•

The recommended wall cutout is ½” larger than the exterior of the pass thru in each direction.

•

Sloped top mounting frames should be attached to the pass thru before installing in wall. If

there is a sloped top on both sides, one side is screwed into the pass thru and the wall, the

opposite side is only screwed into the wall.

•

After installation sealant may be needed to fill any gaps between the pass thru and mounting

frames. Silicone or polyurethane caulk is recommended for this.

Standard Wall Mounted Pass Thru Installation

Before installation locate all mounting frames and mounting frame hardware. The “notched” frame

pieces attach to the vertical sides of the pass thru. Wall mounted pass thrus featuring universal

mounting frames can be installed in a variety of configurations as outlined below.

OPERATION & MAINTINENCE INSTRUCTIONS

MDK18

Pass Thru Air Lock Cleanroom Entry System

4

5

Fire Rated Pass Thru Installation

Most fire rated pass thrus come with pre-installed universal anchoring brackets welded into the

frame throat as shown. These can be anchored to a variety of wall types. Below are the most

common installation methods.

The pass thru must be installed plumb and square to ensure proper door alignment. Shimming may

be necessary to square the cabinet.

Installation Notes

It is recommended that floor

mounted fire rated units be anchored

to the floor in addition to the wall.

Maintain clearance between frame

return flanges and stud for inserting

drywall.

Do not install drywall until you are

certain the frame and pass thru are

plumb and square.

Drywall must extend at least ½” into

the frame.

OPERATION & MAINTINENCE INSTRUCTIONS

MDK18

Pass Thru Air Lock Cleanroom Entry System

6

Operation & Maintenance

After installation the pass thru should be checked for proper function. Door latches and interlock

components should be aligned and operate smoothly.

Interlock Operation

Pass thru interlocks prevent both doors from being open at the same time. When one door opens

the opposite side is locked. The most common interlock styles are listed below.

90° Rotation T-Handles

The T-handles are connected so when one side is unlocked (T-

handle vertical) the opposite side will be locked (T-handle

horizontal). The T-handles can only be turned when both

doors are closed; when one door is open the internal guide

pin is decompressed and will prevent the T-handles from

turning.

Fire Rated / Bio-Design Interlocks

Fire rated and Bio-Design interlocks do not have stainless steel T-handles. They both have interlock

mechanisms which will lock the opposite side automatically as soon as a door is opened.

Magnetic Interlocks

If your pass thru is equipped with magnetic style interlocks the doors are normally locked. The doors

are unlocked by an illuminated capacitive touch button. When both of the doors are closed the

buttons are green the doors can be unlocked. If a door is open the buttons are red and the opposite

door cannot be unlocked.

HEPA filtered units equipped with the optional air shower style controls may also be locked while the

filter is running. Wiring schemtics are included with the manual inside unit.

HEPA Filtered Pass Thru Operation

HEPA filter blowers are set to a velocity of 90ft/min from the factory and feature a speed control for

adjustment in the field. HEPA filters will remove 99.99% of airborne particles at 0.3 µm. Test sheets

and a wiring schematic are included in each pass thru with the manual.

Do not touch the filter media; if the media becomes damaged the filter will need to be replaced.

Air Flow

Recirculating HEPA filtered pass thrus continuously recirculate the air through perforations in the

walls. Single pass HEPA filtered pass thrus purge air through the perforated dirty side door. Air flow

diagrams can be found on our website.

OPERATION & MAINTINENCE INSTRUCTIONS

MDK18

Pass Thru Air Lock Cleanroom Entry System

7

Maintenance

•

At least once a day wipe down interior surfaces and knobs. Contaminate can get into the pass

thru while the dirty side door is open. Particles can float into the pass thru and settle on the

interior. If left uncleaned this contamination can then be transferred into the cleanroom.

•

If the pass thru doors feature polycarbonate or acrylic windows wipe them down lightly with

soft cleanroom wipes to avoid scratching. Verify that any cleaning chemicals used are

compatible the with window material before cleaning.

•

Once a year check that the pass thru is securely mounted. All hardware, including perimeter

mounting screws should be securely fastened.

•

Verify that any cleaning chemicals used are compatible with cabinet materials.

Troubleshooting

Mechanical Interlock T-handles will not turn

Check that the interlock strikes aren’t bent; they should be at a 90° angle to the door.

Mechanical Interlock T-handles turn even when the door is open

Look inside the rectangular slot on the front of the cabinet to check that the guide pin is

functioning properly.