Maintenance

•

At least once a day wipe down interior surfaces and knobs. Contaminate can get into the pass

thru while the dirty side door is open. Particles can float into the pass thru and settle on the

interior. If left uncleaned this contamination can then be transferred into the cleanroom.

•

If the pass thru doors feature polycarbonate or acrylic windows wipe them down lightly with

soft cleanroom wipes to avoid scratching. Verify that any cleaning chemicals used are

compatible the with window material before cleaning.

•

Once a year check that the pass thru is securely mounted. All hardware, including perimeter

mounting screws should be securely fastened.

•

Verify that any cleaning chemicals used are compatible with cabinet materials.

Troubleshooting

Mechanical Interlock T-handles will not turn

Check that the interlock strikes aren’t bent; they should be at a 90° angle to the door.

Mechanical Interlock T-handles turn even when the door is open

Look inside the rectangular slot on the front of the cabinet to check that the guide pin is

functioning properly.

OPERATION & MAINTINENCE INSTRUCTIONS

MDK18

Pass Thru Air Lock Cleanroom Entry System